18 GMC-I Messtechnik GmbH
8 Configuring Device Parameters
General device
parameters can be
configured and
saved with the se-
lector switch in the Setup position.
Select a menu and acknowledge.
Select parameter and acknowledge, change setting and acknowledge.
Limit values ...
Settings x / – = function activated / deactivated
Illumination Background illumination for the LCD. One of three
different conditions can be selected* with the up and
down scroll keys: x: continuously on, –: off
numbers 1 through 9: duration in minutes after which
illumination is automatically deactivated.
Test time Duration of a single test (0 255 s)
Reference voltage: Voltage to which leakage current makes reference
(as a rule nominal line voltage)
Earth fault: During the short-circuit test, testing is also performed
to determine whether or not a connection exists
between L/N and PE (short-circuit to exposed con-
ductive part). We assume that a short-circuit to an
exposed conductive part exists in the event of leak-
age current greater than 15 mA from L/N to PE. This
value should be increased for some DUTs (in particu-
lar high-current consumers), because greater leak-
age currents are present.
Mains wait
Line voltage is first applied to the test socket. However,
testing does not begin until after the duration selected in
seconds under “Mains pause” has elapsed.
* The background illumination of test instruments equipped with panels produced
as of 2014 can no longer be deactivated.
Automode x: for fully automatic test sequences the messeges
are mainly suppressed
Test Sequence ...
Settings x / – = function activated / deactivated
Single-fault
If the single-fault condition has been activated, the test
is interrupted as a failure as soon as an error occurs.
Auto Class PSI Test results (passed or failed) for the various selector
switch positions are automatically assigned to the 8
statistics channels.
inc. Service Error Measurement results are compensated by taking
service error into consideration (measuring error).
IT Network Testing in IT systems can be performed by suppress-
ing tests for U
PE-N
. The U
PE-N
test determines
whether or not voltage is present at PE.
(Leakage current measurement results may other-
wise be distorted.)
Acst Sig, Seq
Acoustic signal is generated for: Incorrect connection of the
DUT, error in the electrical supply system, next test step.
Acst Sig, Meas Acoustic signal is generated for: Measured value
fluctuations, test current polarity reversal
Auto meas. point Prerequisite: Feature KD01*. An acoustic signal indi-
cates whether or not the probe is connected to the
protective conductor. The test sequence is run auto-
matically. Rapid signal frequency: probe connected
to PE; slow signal frequency: next measuring point
Direct Print-Out
Prerequisite: Feature KE01*, see chapter 18 on page 59.
Reports Reports which have been saved to memory can be
selected from a list with an ID number and displayed
(see chapter 18 on page 59).
Select template a report template can be selected for print-out from
of 5 different templates.
SECUSTORE Optimize data transfer for the SECUSTORE adapter
connection (in this configuration, no data can be
saved to the (P)SI module). Neither is a test report
signalled to the RS232 interface.
Service – Time and date settings (if a (P)SI module is used,
the same time and date must also be entered to the
(P)SI menu)
– Service functions (password required)
Setup
18

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST HV wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info