36 GMC-I Messtechnik GmbH
15 Measurements in Accordance with National and
International Standards with Selector Switch in Auto Position
If measurements need to be performed in accordance with given stan-
dards which require specific tests, and if results need to be stored and
documented with a test report, an automatic test sequence is advisable
instead of individual measurements.
Note
+
for tests per EN 60950, EN 61010 and EN 60335
The DUT must be connected to the test socket for the automatic
test sequence! The DUT must comply with safety class I or II.
õ Connect the test instrument to mains power.
A mains connection test is initialized (see chapter 3.2 on page 12).
õ
Connect the DUT to the test socket at the test instrument (see
chapter 7 on page 17). The test instrument initializes
connection type recognition
.
õ Set the selector switch to the Auto position. If the DUT has been con-
nected to the test socket, safety class recognition is performed. Other-
wise, the safety class must be specified manually. In the initial program
window, move the cursor up with the
key into the third line and ac-
knowledge with the key. A safety class can now be selected with
the
and keys and acknowledged with the key.
õ In order to select the regulation for which testing is to be performed,
move the cursor up within the initial program window with the
key
into the top line and acknowledge with the key. The desired stan-
dard can now be selected with the
and keys and acknowledged
with the key. If you always want to test in accordance with a spe-
cific standard, this can be stored as the default standard in the setup
menu. Otherwise, the selected standard only remains active until the
test instrument is disconnected from mains power (factory default set-
ting: VDE 0701-0702).
õ The test sequence can be configured in the Setup menu, limit values
can be changed if necessary and database options can be configured.
õ The test sequence is started by selecting Start test and by acknowledg-
ing with the key (see “Test Sequences”).
Tests which have already been included in chapters 9 through 14 are not
described here again. The only exception is measurement of extension
cables.
15.1 Test Sequences
Test sequences for the various standards are always run in the same
order, assuming that the device under test has been properly connected
and the mains connection test has been passed.
The test sequence can be run step by step with manual activation of each
subsequent step if this function has been specified, or automatically. Step
by step manual operation can be selected in the initial program window in
the setup menu under “Sequence” ... “manual sequence”.
Safety class: SC I or SC II is automatically recognized by the test instru-
ment. If the safety class is not correctly recognized, you can change it
manually. This manual correction is not stored however.
Visual inspection: If “visual inspection” has been activated
in the initial program window under “Sequence”.
If a part is recognized as defective by the user, it must be identified as
such by selecting it with the cursor and acknowledging with the key.
Protective conductor resistance measurement
(for SC I devices under test only)
Note
+
This test cannot be skipped. Exception:
This measurement can be skipped by pressing
(when mes-
sage „Connect probe to protective conductor“ is displayed) if a
protective conductor connection is not possible.
Evaluation of protective conductor testing
High-voltage test for DIN VDE 0701 Part 1, Appendix E, EN 60950,
EN 61010, EN 60335, EN 60601, if selected in the setup menu.
Insulation resistance measurement
IEC 62353: only if selected in the setup menu of the
initial program window under “Sequence ...
EN 60601: only if selected in the initial program window
under „Test Conditions“
DIN VDE 0701-0702 only if R-INS is activated in the initial program
window
Evaluation of insulation test
Leakage current measurement (various single-fault conditions are run de-
pending on classification)
Evaluation of each individual leakage current measurement, see also
chapter 21.2
Evaluation of the overall test
36

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST HV wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info