590775
38
Zoom out
Zoom in
Vorherige Seite
1/72
Nächste Seite
38 GMC-I Messtechnik GmbH
All of the possible sequence settings for all of the regulations are listed
below.
Select the Setup... menu from the initial program window and
acknowledge.
Sequence 1 (repair tests, periodic testing)
store All of the settings in the setup menu, i.e. the configu-
ration of the measuring parameters and the current
limit values can be saved with this command for the
selected test standard. These values remain active
even after setting the selector switch to a different
position, and after disconnection from mains power.
Sequence See page 39.
Limit Values See chapter 4.6 on page 14.
Database Start with ID no.
x: Before each measurement is started, an entry
prompt appears requesting entry of an ID no.
An individual number can be entered (max. 20
characters) with the keypad at the (P)SI module
(optional), read in with a barcode scanner
(optional) or directly selected from a list.
If an incorrect entry is made:
Only complete lines can be deleted, and deletion
is only possible with the key at the instrument.
ID no. = test sequence (Feature KB01 or
SECULIFE ST HV
).
See chapter 16 on page 58.
Additional Parameters
R-PE with clip x: Protective conductor resistance can be
determined with the help of the WZ12C current clip.
High-voltage x: A high-voltage test is conducted
(prerequisite: Feature F02 or SECULIFE ST
HV)
Sequence 2 (type tests)
Type: ........ Sequences 1 and 2 differ primarily as regards the
type parameter. The user can set up any desired type
of device under test with sequence 2, for which the
same limit values and measuring parameters always
apply. Up to 125 different types can be defined. DUT
types are predefined for sequence 1 (EN 60601).
Designations made up of alphanumeric characters
(max. 10 characters) can be entered at the keypad of
a (P)SI module, or at a PC with a terminal program.
The test sequence, including all limit values, is saved
together with the selected type. We recommend
using the characters which identify the respective
standard as the initial characters in the designation. If
a type is queried from the initial program window
which does not belong to the currently selected stan-
dard, these characters identify the test sequence as
belonging to another standard.
Limit Values See chapter 4.6 on page 14.
Sequence See page 39.
Store as
Default All of the settings in the setup menu, i.e. the configu-
ration of the measuring parameters and the current
limit values can be saved with this command.
These values remain active even after setting the
selector switch to a different position, and after dis-
connection from mains power.
Delete The type which has been selected in the initial pro-
gram window can be deleted.
38

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST HV wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info