15.3 Configuring Measuring Parameters
Depending upon the test regulation, various measuring parameters can
be configured for the test sequence (settings: x / – = function activated /
deactivated). All of the possible parameters for all of the regulations are
listed below. The Sequence... menu is accessed via the setup parameter in
the initial program window for the respective regulation.
Select the Sequence... menu and acknowledge.
Select a parameter, acknowledge, change and acknowledge the change.
Visual inspection This menu appears at the very beginning of the test
sequence.
Manual sequence Each test step must be acknowledged with the
key (see test sequence in chapter 8 on page 18
regarding test duration for automatic sequence).
Autostore After testing is completed, test data are automatically
stored to the test instrument(Feature KB01), or in the
(P)SI module (accessory).
Mains polarity
reversal L an N are reversed each time line voltage is applied
to the test socket.
Classification Questions regarding classification appear if limit
values are exceeded, see chapter 5 on page 15.
SC III U
V
Supply voltage is measured instead of insulation
resistance for active devices under test.
R-INS LN-PE Insulation resistance measurement is performed
between phase and neutral conductors, and the
protective conductor.
R-INS AP-PE Insulation resistance measurement is performed from
an application part to the protective conductor.
HV test A high-voltage test is performed
(prerequisite: Feature F02 or SECULIFE ST
HV)
HV test duration Duration of a high-voltage test (5 ... 60 s)
Auto (test) method The instrument recognizes whether or not the device
under test can be switched on:
Leakage or residual current, or insulation resistance
and earth leakage current are measured accordingly.
R-PE AC > 10 A
Protective conductor measurement with 200 mA DC
,
10 A AC
(feature G00 or
SECULIFE ST)
oder 25 A AC
(
feature
G01 or
SECULIFE ST HV)
R-PE with clip Protective conductor resistance can be determined
with the help of the WZ12C current clip
No I
HL
for SCI Contact current is not tested for SCI.
Mains wait A waiting period between switching mains power on
and starting the test can be entered here, e.g. in
order to suppress the display of recorded values
which occur during warm-up of the device under
test.
Patient auxiliary
current Measurement is performed with patient auxiliary
current
Adapter for socket Limit values are activated for permanently connected
devices. A device under test which is normally per-
manently installed can be connected to the test
socket via an adapter. No voltage may be applied to the
test socket when this test method is used.
First measured
values A menu for the entry of the first measured value
appears during the test sequence.
39

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Gossen Metrawatt SECULIFE ST HV wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info