Select connection, select test regulation
VDE 0701 part 1
, classify DUT (SC I, II or III),
part 1: extension cable X/– (with/without), Appendix E: HV test X/– (with/without)
Visual inspection
O.K.?
SC I: RPE
OK?
I
L
OK?
Short-circuit
at DUT?
Test
OK?
Yes/No
No
Yes
Display results, save/print report
Yes
No
Yes
No
No
Test
O.K.?
Framed with dashed line:
The test is only run if it has been activated in
the initial program window, or in the Setup
menu under Sequence...
Switch line voltage to test socket, start function test
Start
function test
Safety Class I only: Contact all exposed, conductive parts with the
probe. The test can be repeated as often as desired for various pro-
tective conductor parts with the manual sequence*.
* If it is not clear whether or not all ex-
posed, conductive parts are connect-
ed to one another or to the protective
conductor, testing can be performed
in the manual mode.
Switch DUT on
RINS+I
EL
OK?
or
Yes
No
No
Yes
Yes
Test HV? HV OK?
Apply high voltage
Yes
No
Yes/No