Aborting a measurement
You can stop the measurement at
any time by pressing the ON/OFF
button. The BP monitor deflates the
cuff instantly.
The memory function
The SC 7100 saves up to 50
measurements in every memory area.
Together with each measurement
(systolic, diastolic and heart rate), the
date and time are also saved. The
measurement with the longest residence
time in the memory is assigned number 1.
All other measurements increment by one
until number 50. Then the memory is full.
The next measurement occupies memory
location 50. The measurement formerly at
location 50 changes to location 49 (etc.,
4847 .... 21). The oldest
measurement formerly at location 1 is
lost.
Delete a blood pressure value from the
memory
Press the MEMORY key to view the blood
pressure value you want to delete.
Pressing the ON/Off button instantly (and
holding it for at least 4 secs.) displays dEL
(for "delete") and
the number of the
memory location.
Press the ON/Off button instantly and the
value will be deleted.
Delete all blood pressure values in the
memory
Press the MEMORY key to display any
blood pressure value. Pressing the
ON/Off button instantly (and holding it for
at least 4 secs.) displays
dEL (for "delete") and the number of
the memory location:
Press the MEMORY key instantly and the
LCD displays
dEL (for "delete") and ALL (for all
memory locations).
Press the ON/Off button instantly and all
values will be deleted. Time and date are
not deleted.
- 6 -
WHO Interpretation of blood pressure
readings
29/12/2007
Meaning Systolic
/ mmHg
Diastolic /
mmHg
optimal < 120 < 80
normal < 130 < 85
high-normal
130 -
139
85 - 89
mildly
hypertonic
140 -
159
90 - 99
hypertonic,
medium
severity
160 -
179
100 - 109
severely
hypertonic
> 180 > 110
The World Health Organization (WHO)
recommends medical treatment, when the
blood pressure value exceeds 140/90
mmHg.
Examples:
Cleaning and maintenance
Do not expose the BP monitor to
extreme temperature, moisture, dust or
direct sunlight.
The cuff contains a sensitive airtight
bulb. Treat it with care and avoid any
stress by torsion or kinking.
Clean the BP monitor with a soft, dry
cloth. Do not use fuel, thinner or other
solvent. Remove specks on the cuff
carefully with a wet cloth and soap water.
Do not wash the cuff.
Do not drop the BP monitor or treat it
with force.
Do not try to open the BP monitor. This
will make the factory calibration invalid
- 7 -
Troubleshooting and remedies
If a fault is detected during a
measurement, the BM monitor displays
„Err“ and a symbol (code for service
personnel).
From many years of experience we know
that most faults are the consequence of
wrong use. The most frequent causes of
faults are excessive movement during the
measurement, wrong application of the
cuff, use of wrong batteries or use of a
rechargeable battery.
Message
Error
description
Correction
The pressure
measured was
lower than
20mmHg.
Please measure
again.
The pressure
measured was
higher than
300mmHg.
Please measure
again.
Pumping error.
Please check cuff
and try again.
The pressure
can not be
measured due
to signal noise.
Please measure
again.
The pumping
pressure is
higher than
300mmHg.
Please measure
again.
or no
display
Low batteries.
Please check and
replace the
batteries if
necessary.
Wait at least 5 minutes before repeating
the measurement.
Periodic re-calibration
We recommend periodic checks of the
pressure indication every 2 years.
- 8 -
Referenced norms
For registration, a defined number of
blood pressure monitors was subjected to
stringent clinical testing. These monitors
used the same program for blood
pressure measurement as your monitor
uses. Thus, this program is also regarded
as clinically tested.
The manufacture of the blood pressure
monitor complies with the European
norms for blood pressure monitors: EN
60601-1, EN 1060-1 to EN 1060-3.
Electromagnetic compatibility:
The blood pressure monitor complies with
the requirements of European norm EN
60601-1-2.
The requirements of EU directive
93/42/EEC concerning medical devices,
class IIa are complied with.
Warranty
We grant a warranty of two years for
blood pressure monitor SC 7100. Within
this period, defects of the BP monitor will
be repaired free of charge. You can also
have your BP monitor repaired by our
customer service after the warranty
period. We will provide you with a free
cost estimate for the repair.
The warranty does not cover damage due
to improper handling, accident, natural
wear and tear (cuff), non-compliance with
the instructions for use or unauthorized
modifications of the monitor.
Excluded from the warranty are:
the packaging, batteries and the
instructions for use.
Disposal note
The disposal of batteries is regulated.
Spent batteries are collected at collection
points provided by battery dealers and the
communities.
Dispose of the blood pressure monitor as
prescribed. i.e., at a collection point for
electrical scrap.
- 9 -
Specifications
Measuring method: Oscillometric
Measuring range:
Pressure: 20 to 280 mmHg
Heart rate: 40 to 195 / min
Measuring accuracy:
Stat. pressure: +/- 3 mmHg
Heart rate: +/- 5 %
Measuring value memory: 2 memories
each with 50 locations for systolic /
diastolic / heart rate, date and time
Display: LC display
Pressure pick-up: Semi-conductor
pressure sensor
Automatic switch-off: after approx. 3
minutes of inactivity
Voltage supply: 2 x LR03 (alkali
batteries),
Working temperature range:
+10°C - +40°C
Storage temperature range:
- 10°C - +50°C
Cuff length: for wrist circumference from
135 mm to 195 mm
Specifications may change as technical
improvements take place
SCALA Electronic GmbH
Ruhlsdorfer Straße 95
D-14532 Stahnsdorf, Germany
Service firm SES GmbH
Ruhlsdorfer Straße 95
D-14532 Stahnsdorf,
Germany
Rev. 18/04/2011/GB
- 10 -
4

Brauchen Sie Hilfe? Stellen Sie Ihre Frage.

Forenregeln

Missbrauch melden von Frage und/oder Antwort

Libble nimmt den Missbrauch seiner Dienste sehr ernst. Wir setzen uns dafür ein, derartige Missbrauchsfälle gemäß den Gesetzen Ihres Heimatlandes zu behandeln. Wenn Sie eine Meldung übermitteln, überprüfen wir Ihre Informationen und ergreifen entsprechende Maßnahmen. Wir melden uns nur dann wieder bei Ihnen, wenn wir weitere Einzelheiten wissen müssen oder weitere Informationen für Sie haben.

Art des Missbrauchs:

Zum Beispiel antisemitische Inhalte, rassistische Inhalte oder Material, das zu einer Gewalttat führen könnte.

Beispielsweise eine Kreditkartennummer, persönliche Identifikationsnummer oder unveröffentlichte Privatadresse. Beachten Sie, dass E-Mail-Adressen und der vollständige Name nicht als private Informationen angesehen werden.

Forenregeln

Um zu sinnvolle Fragen zu kommen halten Sie sich bitte an folgende Spielregeln:

Neu registrieren

Registrieren auf E - Mails für Scala SC7100 wenn:


Sie erhalten eine E-Mail, um sich für eine oder beide Optionen anzumelden.


Das Handbuch wird per E-Mail gesendet. Überprüfen Sie ihre E-Mail.

Wenn Sie innerhalb von 15 Minuten keine E-Mail mit dem Handbuch erhalten haben, kann es sein, dass Sie eine falsche E-Mail-Adresse eingegeben haben oder dass Ihr ISP eine maximale Größe eingestellt hat, um E-Mails zu erhalten, die kleiner als die Größe des Handbuchs sind.

Ihre Frage wurde zu diesem Forum hinzugefügt

Möchten Sie eine E-Mail erhalten, wenn neue Antworten und Fragen veröffentlicht werden? Geben Sie bitte Ihre Email-Adresse ein.



Info